A simple method to achieve results similar to the freeze-etching technique of Moor et al. (1961) is described. The frozen tissue is cut under liquid nitrogen with a razor blade outside the evaporator rather than inside with a cooled microtome. The conditions of the experiment do not favor sublimation, and it is proposed that the structure of the replica be explained by local faults in the cleavage plane which leaves structures, such as membranes, standing above the ice. Micrographs of replicas of glycerol-protected frozen small intestine of mouse prepared by the method are presented and the structural details they show are discussed. The problem of vapor-deposited contamination is discussed. It is concluded that this is a practical method for obtaining electron micrographs that are relatively free of artifact, and that further improvements may be expected from the use of rapidly frozen fresh tissue and a clean vacuum system, possibly of the ion-pumped type.
Skip Nav Destination
Article navigation
1 June 1966
Article|
June 01 1966
A SIMPLE FREEZE-FRACTURE REPLICATION METHOD FOR ELECTRON MICROSCOPY
Stanley Bullivant,
Stanley Bullivant
From the Mixter Laboratories for Electron Microscopy and Neurochemistry, Neurosurgical Service, Massachusetts General Hospital, Boston
Search for other works by this author on:
Adelbert Ames, 3rd
Adelbert Ames, 3rd
From the Mixter Laboratories for Electron Microscopy and Neurochemistry, Neurosurgical Service, Massachusetts General Hospital, Boston
Search for other works by this author on:
Stanley Bullivant
From the Mixter Laboratories for Electron Microscopy and Neurochemistry, Neurosurgical Service, Massachusetts General Hospital, Boston
Adelbert Ames, 3rd
From the Mixter Laboratories for Electron Microscopy and Neurochemistry, Neurosurgical Service, Massachusetts General Hospital, Boston
Received:
January 27 1966
Online ISSN: 1540-8140
Print ISSN: 0021-9525
1966
J Cell Biol (1966) 29 (3): 435–447.
Article history
Received:
January 27 1966
Citation
Stanley Bullivant, Adelbert Ames; A SIMPLE FREEZE-FRACTURE REPLICATION METHOD FOR ELECTRON MICROSCOPY . J Cell Biol 1 June 1966; 29 (3): 435–447. doi: https://doi.org/10.1083/jcb.29.3.435
Download citation file:
Sign in
Don't already have an account? Register
Client Account
You could not be signed in. Please check your email address / username and password and try again.
Could not validate captcha. Please try again.
Sign in via your Institution
Sign in via your InstitutionEmail alerts
Advertisement
Advertisement