Table 3.

Values of model parameters that best simulated photocurrents in dark-adapted bass single cones: Response to 10-msec light flashes

 Parameters Categorya Units Cone 1 Cone 2 Cone 3 Mean ± SDb 
VP* Intensity Statistical VP* 167 17,443 174 17,136 173 7,720 Fig. 4  
 γmax Adjustable µM/s 100c 65 100 70 105 68  
 μ0 Invariant 1/s 0.5 0.5 0.5 0.5 0.5 0.5 — 
 ωλ Invariant  0.1 0.1 0.1 0.1 0.1 0.1 — 
T* Ψ0 Adjustable  230d 230d 215 215 230 230 231 ± 19 
 ωact Invariant  0.69e 0.69e 0.69e 0.69e 0.69e 0.69e — 
PDE* βsub Adjustable µM/s per molecule 0.185 0.185 0.323 0.323 0.416 0.416 0.334 ± 0.091f 
 αPDE Adjustable 1/s 17g 47 19 14 28 Fig. 4  
CNG ion channel Idark Statistical pA 22.2 22.2 18.8 18.8 42.3 42.3 27.3 ± 10.5 
Ca2+ buffer KHA Adjustable µM 0.030 0.030 0.044 0.044 0.091 0.091 0.052 ± 0.025 
 CHA Adjustable µM 21.4 21.4 4.6 4.6 66 66 20.9 ± 12.9 
 B Adjustable  10.6 10.6 15 15 11.5 ± 3.9 
Ca2+ efflux JCamax Statistical pA 3.23 3.23 7.63 7.63 4.87 ±1.88h 
 KCaexc Adjustable µM 0.025 0.025 0.005 0.005 0.025 0.025 0.019 ± 0.009h 
 Parameters Categorya Units Cone 1 Cone 2 Cone 3 Mean ± SDb 
VP* Intensity Statistical VP* 167 17,443 174 17,136 173 7,720 Fig. 4  
 γmax Adjustable µM/s 100c 65 100 70 105 68  
 μ0 Invariant 1/s 0.5 0.5 0.5 0.5 0.5 0.5 — 
 ωλ Invariant  0.1 0.1 0.1 0.1 0.1 0.1 — 
T* Ψ0 Adjustable  230d 230d 215 215 230 230 231 ± 19 
 ωact Invariant  0.69e 0.69e 0.69e 0.69e 0.69e 0.69e — 
PDE* βsub Adjustable µM/s per molecule 0.185 0.185 0.323 0.323 0.416 0.416 0.334 ± 0.091f 
 αPDE Adjustable 1/s 17g 47 19 14 28 Fig. 4  
CNG ion channel Idark Statistical pA 22.2 22.2 18.8 18.8 42.3 42.3 27.3 ± 10.5 
Ca2+ buffer KHA Adjustable µM 0.030 0.030 0.044 0.044 0.091 0.091 0.052 ± 0.025 
 CHA Adjustable µM 21.4 21.4 4.6 4.6 66 66 20.9 ± 12.9 
 B Adjustable  10.6 10.6 15 15 11.5 ± 3.9 
Ca2+ efflux JCamax Statistical pA 3.23 3.23 7.63 7.63 4.87 ±1.88h 
 KCaexc Adjustable µM 0.025 0.025 0.005 0.005 0.025 0.025 0.019 ± 0.009h 
a

Values of invariant parameters first used to compute dark current are not repeated here. They have the same values listed in Table 2.

b

Mean ± SD; n = 18 single cones.

c

γmax experimental value is 100–150 s−1 (Tachibanaki et al., 2005).

d

This value was initially constrained to 150 s−1 from experimental data (Leskov et al., 2000).

e

Experimental data (Gibson et al., 2000).

f

Experimental values of PDE* kcat range from 2.2 to 8 × 103 s−1 (Dumke et al., 1994; D’Amours and Cote, 1999; Muradov et al., 2009). From the definition of βsub (Eq. 3.11) and the highest kcat value, the 0.334 value indicates the effective intracellular volume for cGMP is, on average, 0.08 pL, rather than the 0.19 pL used throughout these calculations. This factor of approximately two can probably be accounted for by the fact that the volume, in fact, changes some from cell to cell and is not the rigid assigned value of 0.19 pL. There is experimental uncertainty because the size of each and every cone recorded from was not measured. Also, the actual concentration of cGMP may be a little smaller than the computed value because there is intracellular cGMP binding (Olson and Pugh, 1993), a fact that would be revealed as a smaller apparent volume.

g

This value was initially constrained to 10 s−1 at dim lights from the analysis of dark current noise in bass single cones (Holcman and Korenbrot, 2005).

h

With these values, the computed time constant of Ca2+ clearance from bass cone outer segment is 40.0 ± 15.4 msec, similar to the experimental value of 43 ± 9.8 msec in the tiger salamander cone (Sampath et al, 1999).

or Create an Account

Close Modal
Close Modal